Highly accurate temperature control is possible using microcomputer PID control.
The display uses a liquid crystal screen with excellent visibility.
Equipped with 4 types of timer function (ON/OFF/ON/OFF/SET) as standard, and the program function can input temperature and time up to 8 steps.
With dedicated software (CDROM) and communication cable (optional) for data recording and communication, you can easily manage data on your PC.
The sample temperature can be directly controlled using an external temperature sensor (optional).
Equipped with an alarm signal output as standard, it is possible to stop other equipment and start the backup system in the event of an error.
A safety lamp is provided between the main unit front operating section and the hot plate section (lights up until the top plate temperature drops below 55°C).
The safety device uses a double lock mechanism consisting of two independent circuits.
The temperature rises to the maximum in approximately 30 minutes.
Top plate temperature distribution width is highly accurate at 4°C (at 200°C setting).
Specification
Maximum temperature (°C): 300
Plate Size (mm): 400 x 300
Temperature Setting Range: Room Temperature +10~300°C
Temperature Setting: Digital Input Type
Temperature Distribution Width: 4°C(The temperature distribution width is a reference value for an indoor environment of 25°C(No wind, no load, internal sensor used) and 200°C.)
Temperature control system: microcomputer PID control system, ON/OFF control system
Temperature Display Method: LCD Digital Display (Setting/Measurement Temperature Simultaneous Display/Heater Output Bar Graph Display)
Internal Temperature Sensor: K Thermocouple
Safety Devices: Self-Diagnosis Function (Upper and Lower Limit Temperature Abnormality Detection, Heater Abnormality Detection, Sensor Abnormality Detection), Abnormal Heating Prevention Device (Thermostat), Overcurrent Fuse