The display part is large and easy to see, and it is a small and lightweight type that is easy to grip.
SN3200-T01N sensor (attached) A wide range of temperature measurements can be made for robots such as surface temperature measurement of iron plates and roads, temperature measurement of asphalt, temperature measurement of high-temperature oil and liquid, temperature measurement of showcases, temperature measurement of kettles, etc., by combining with sensors other than those in the set.
It also has a temperature compensation function, so important temperature control is possible.
Tressor calibration is also possible in combination with most optional sensors. (Expenses not included)
Specification
Set with SN-3200II-00+SN-3200-T01N (T-sensor)
Accuracy Guaranteed Temperature Range: -30.0°C to +100°C
Accessories: AA batteries, instruction manual, neck strap
Body Specifications
▼ Input Type: Thermocouple (ST, SK)
▼ Mass (g): Approx. 180 (including battery)
▼ Power: 2 AA batteries
▼ Input point: 1 point
▼ Other functions: Calculation functions: Maximum/minimum value/variation, hold, battery alarm display, sensor type discrimination, temperature correction function
▼ Operating ambient temperature: 0~ 50°C
▼ Ambient humidity: 80%RH or less (no dew condensation)
▼ Display resolution (automatic) 0.1°C (decimal point display): -40.0°C to +199.9°C, 1°C (integer display): 200°C or more
▼ Body Display Capacity (°C) ST: -99.9°C to +350°C, SK: -99.9°C to +1250°C (depending on the sensor used)
[Body accuracy] ± (0.1%rdg+0.3°C): -30.0°C to +100.0°C, ± (0.2%rdg+1°C): +100.0°C to +800°C (* Ambient temperature 23°C±5°C)
[Sensor Accuracy] (ST Sensor) ±0.5°C: -30.0 to +100.0°C, (SK Sensor) ±1.5°C (Class 1): 0~375°C, ± (0.004 x ItI) °C: +375°C to +800°C, (Surface Sensor) ±2.5°C (Class 2): 0°C to +333°C, ± (0.0075 x ItI) °C: +333°C to +1200°C
[Overall accuracy] Total body accuracy and sensor accuracy: Example 1. Using T01N sensor: 100°C×0.1% when measured at 100°C =0.4°C (body accuracy) +0.5°C (sensor accuracy) =0.9°C (overall accuracy ±0.9°C) Example 2. Using K01 sensor: 100°C×0.1% when measured at 100°C =0.4°C (body accuracy) +1.5°C (sensor accuracy) =1.9°C (overall accuracy ±1.9°C)