Air Jacket Natural convection formula Temperature Control Range: R.T. + 5 to 60 ° C Temperature Adjustment Accuracy: ± 0.5 ° C (At 37 ° C) Temperature Distribution Accuracy: ± 3.5 ° C (At 37 ° C) Temperature distribution (at 37 degrees C): Spatial temperature deviation 2.3 ° C, Temperature gradient 3.4 ° C Timer time setting range: 0 to 99 hours 59 minutes Temperature Control/Setting/Display: P.I.D Control SSR Output/Sheet Key Input/Digital Display (Temperature Display Correction, Auto Start, Auto Stop 1 and 2, Power Failure Recovery Setting) Safety Function: Alarm Function (Upper limit temperature, door open, sensor, power failure, temperature uncontrollable, watchdog), Independent Excess Temperature Rise Prevention System (Fixed at 70 ° C), Fuse Heater: 100 W temperature sensors: thermistors Interior, Inner Door: SUS304 inner dimension (Width x Depth x Height) : 310 x 240 x 220 mm Inner volume (L): about 16 shelf board Load tolerance/material: Equal Distribution Max. 3kg/Sheet/PE Coating Number of shelf board Supplies/Shelf pitch/number of step: 2 pieces/60 mm/3 stages Operating ambient temperature range: 5 to 35 ° C external dimension (Width x Depth x Height): 395 x 355 (375) x 420 mm/* The Dimensions in () includes bumps. Weight (kg) : about 14 power supply Input/power supply Voltage : 1.5 A 150 VA/AC100V 50/60 Hz Code Number: 197980 *Performance measured at R.T. 25 ° C, rating power supply Voltage, 50Hz, and no load. *There may be a difference between the set temperature and the actual temperature due to the characteristics of the product. *When the shaker is inserted into the chamber, the temperature control range is + 10 ° C or higher. Also, inside temperature may fluctuate. *Performance description of temperature distribution conforms to JIS C 60068 -3 -5 JTM K07 -2007 (Japan Testing Machine Industry Association, performance testing method and performance display method for temperature testing tank). *power supply input does not include internal outlet capacity.