Normal maximum temperature is 1200°C and the temperature is increased rapidly.
Using a core tube and gas replacement unit allows heating in an inert environment.
You can choose the material of the core tube according to the application.
In addition to the automatic overtemperature prevention function linked to the set temperature, it also has a completely independent overtemperature prevention function and a burnout function to detect sensor abnormalities.
It has a safety function to cut off the power to the heater when the door is opened.
Specifications
Model number: TMF-700N
Control system: PID control, SSR output
Inner furnace dimension (mm): 70 x 300
Temperature settings range: 100 - 1200℃
Sensor: R thermocouple x 2
Function: Program setting (16 steps, 8 patterns), timer
Safety device: Independent excessive rise prevention function (set temperature digital input), door switch, electrical leakage breaker, self-diagnosis function (main relay is OFF when sensor is disconnected or heater is disconnected)